The following features have been added in the Release 2.15.0.
- Added Munsell images
- Gateware files can now be read from PCMCIA/CF card when
booting
- HDCP Compliance Test results now written to HTML file
after each test is finished.
The following anomalies have been corrected in the Release
2.15.0.
- Removed HDMI PCP selection.
- Eliminated video noise.
- DisplayPort HDCP sink Compliance Test now reflects current
link rate.
- Fixed Box250mm's image not available message for DMT0660.
- DisplayPort bit rate and lane settings show even if link
training fails.
- DisplayPort low-level settings are now independent of
link training status.
- CTS version number added to report headers.
- Can now send idle pattern and disable video on DisplayPort
while link is up.
- Show clock regeneration M and N values.
- Can now retrain link on command.
- Added link training debug support.
- Added DPTX:EDID? command to read EDID from a malfunctioning
sink.
- Hot plug pulse restarts authentication in HdcpProd image.
- Improved hot plug detection.
- Fixed R/G/B component gating.
- YCbCr to YYY switch now produces correct colors.
Note: The release files can only be applied
to the 882E-DP DisplayPort generator (not the 882CA/882C/881C, 882D, 882E HDMI generator or 882EA HDMI generator).
The Release 2.15.0 is comprised of the following software components.
| Software Component |
Version |
Change Status |
| Boot Rom |
01.04.11 |
no change |
| vxWorks |
20.1885600 |
new with 2.15.0 |
| Gateware 1 |
451C,8,2202008 |
new with 2.15.0 |
| Gateware 2 |
451A,15,11162007 |
no change |
| Gateware 3 |
451B,14,11162007 |
no change |
Java JRE required |
1.5.06 |
no change |
Now you can use Quantum Data's new AutoUpdate utility to upgrade
your generator firmware and gateware. The AutoUpdate utility automates the process of upgrading firmware. It guides the user through the upgrade process, checking the Quantum Data website for the latest version and installing all files in the correct location on the generator.
Read the AutoUpdate Instructions.
Download AutoUpdate Version 1.1.1 for 880 series (includes instructions)
You can upgrade your generator using the procedures in the Users
Guide. Please note that you will have to verify that your
current generator is operating with Release 2.14.0. To verify that your current
generator is operating with this release, enter the
following command:
verf?
Release 2.14.0 should return the following information:
20.1885400, 01.04.11
After verifying that the generator is operating with
Release 2.14.0, upgrade your generator using the procedures
below:
- Download the 2.15.0 Zip file to your PC and unzip the
file. There will be a System and a Library directory.
- Upgrade the generator using "Upgrading without using PC Card" procedure at this link.
- After power cycling the generator, verify that the generator
is running the current 2.15.0 version of the vxWorks file
by entering the following command:
verf?
20.1885600, 01.04.11
- Verify that the vxWorks file is: 20.1885600.
- Verify that the BootRom version is: 01.04.11.
- Verify that the generator is running the current 2.15.0
version of the gateware files by entering the following
command:
verg?
451C,8,2202008:451A,15,11162007:451B,14,11162007
- The gatewares are separated by colons. Verify that the
gateware files are: 451C,8,2202008:451A,15,11162007:451B,14,11162007.
Whether you use AutoUpdate or the manual upgrade method, you must update the OCM firmware in the DisplayPort transmitter
and receiver chips:
- Download DisplayPort OCM firmware ZIP file DP_OCM_2_15_0.zip
- Uzip the files to your PC.
- Use FTP Browser to copy files SafeLite_ISP_I2C_BD_Driver.hex and gm60028BD.hex to the generator's System directory.
- At a command prompt issue the command DPTX:PROG 0.
- This
may take several minutes. Watch display and command prompt.
- When complete, power cycle the generator.
- Using FTP Browser, delete file gm60028BD.hex from the generator
- If your generator has a DisplayPort Receive Port, continue with these steps:
- Use FTP Browser to copy file gm68020BD.hex to the generator's System directory.
- At a command prompt issue the command DPRX:PROG 0
- This
may take several minutes. Watch display and command prompt.
- When complete, power cycle the generator.
- Using FTP Browser, delete file gm68020BD.hex from the generator
|  |