Feature Enhancements for 882EA Release 2.10.5 - vxWorks: 20.1884605

Known Issue: HDCP Compliance Test is broken for source and repeater tests. We are working on a new gateware that fixes the problem. If you cannot wait for the fix, please use release 2.10.2 instead.


There are no new features for the 882EA included in the Release 2.10.5.

The following anomalies have been corrected in the Release 2.10.5.

  • Fixed the problem where HdcpProd or Hdcp2 image does not react to hot plugs correctly.
  • Fixed the problem where CEC bit corruption wasn't working. CEC Compliance Test 9.5-1 can be run properly now.
  • Reworded EDID Compliance Test per HDMI CTS 1.3b. "Too many preferred" means the same as "Native DTD count larger than number of DTDs."
  • Fixed the problem where the generator hangs when leaving HdcpProd in certain cases.
  • Fixed the problem where the web interface's HDCP Compliance Test GUI applet couldn't launch.
  • General HDCP Compliance Test improvements.
HDCP Compliance Test notes (only applicable to 882EA with HDCP Compliance Test option)
Test coverage limitations of the HDCP Compliance Test

  • Test steps TT01 and TT02 of Test Section 1A-01 are not tested because the application does not have visability into the control bits of the TMDS data islands. Therefore a Pass result on the test report only applies to the tests within that section that are executed.

  • Test steps TT01 and TT02 of Test Section 1B-01 are not tested because the application does not have visability into the control bits of the TMDS data islands. Therefore a Pass result on the test report only applies to the tests within that section that are executed.

  • Test Section 1A-08 is not tested because this test requires a DVD with a list of KSVs that have been revoked.

  • Test Section 1A-09 cannot be tested because the application cannot collect the TMDS video.

  • Test Section 2C-01 test S302 cannot be tested because the application cannot currently process Pj. Therefore a Pass result on the test report only applies to the tests within that section that are executed.

  • Test Section 2C-04 cannot be tested because the application cannot collect TMDS data.


Applicable to all tests

  • do not use CMD Terminal tool in the generator's web interface to run the Compliance Test via command line; use Telnet (Ethernet) or HyperTerminal (serial) instead

  • try running the tests in batch mode by using the front panel interface

  • if there is a problem, try running the tests one by one via the command line (see the repeater AB test procedure below as an example and refer to the user guide)

  • running the tests using release 2.6.1 (20.1883610 firmware) saves the test reports in /tffs0/library/reports/

  • running the tests using a beta firmware (20.1883611 and later) or release 2.10.2 (20.1884602 firmware) saves the test reports in /card0/library/reports if there is a PCMCIA card inserted in the 882CA/882EA (/tffs0/library/reports/ if no PCMCIA is inserted)

Applicable to the sink tests

  • you can monitor the transactions on Tx port 1 using the ACA

Applicable to the source tests

  • you can monitor the transactions if you have another 882C/882CA/882E/882EA and a TPA-ACA-3R

  • connect the output of the source device to TPA-ACA-3R's DUT 1, the input 1 of 882CA/EA (test equipment) to TPA-ACA-3R's DUT 2, and output 1 of a different 882C/882CA/882E/882EA (ACA monitoring device) to TPA-ACA-3R's OUT 1

  • use the ACA on the 882C/882CA/882E/882EA that is not the test equipment to monitor the transactions on Tx port 1

Applicable to the repeater AB tests

  • use a separate source device and connect it to the upstream port of the repeater

  • use the 882CA/EA's Rx 1 port as the sink device and connect it to the downstream port of the repeater

  • you can monitor either downstream or upstream transactions (not both at the same time) if you have another 882C/882CA/882E/882EA and a TPA-ACA-3R

  • in order to monitor the downstream transactions, connect the downstream of the repeater device to TPA-ACA-3R's DUT 1, the input 1 of 882CA/882EA (test equipment) to TPA-ACA-3R's DUT 2, and output 1 of a different 882C/882CA/882E/882EA (ACA monitoring device) to TPA-ACA-3R's OUT 1

  • in order to monitor the upstream transactions, connect the output of the source device to TPA-ACA-3R's DUT 1, the upstream of the repeater device to TPA-ACA-3R's DUT 2, and output 1 of a different 882C/882CA/882E/882EA (ACA monitoring device) to TPA-ACA-3R's OUT 1

  • use the ACA on the 882C/882CA/882E/882EA that is not the test equipment to monitor the transactions on Tx port 1

Applicable to the repeater C tests

  • connect the 882CA/EA's Tx 1 port to the upstream port of the repeater


  • connect the 882CA/EA's Rx 1 port to the downstream port of the repeater


  • make sure HdcpProd image is not selected on the 882CA/882EA before running the tests


  • you can monitor the downstream transactions on Rx port 1 using the ACA


  • you can monitor the upstream transactions on Tx port 1 using the ACA
Sink test procedure via the command line interface

    CPTX:DUTT 0 // set up for source test
    CPTX:CPTR 16 // configure 2C-1 test
    CPTX:CPTU // start the test
    // save the test report to the PC and rename it

    // repeat above steps for test numbers 17 and 18

    CPTX:CPTR 19 // configure 2C-4 test
    CPTX:CPTU // start the test
    // save the test report to the PC and rename it
Source test procedure via the command line interface

    CPTX:DUTT 1 // set up for source test
    CPTX:CPTR 1 // configure 1A-01 test
    CPTX:CPTU // start the test
    // save the test report to the PC and rename it

    // repeat above steps for test numbers 2 to 14

    CPTX:CPTR 15 // configure 1B_06 test
    CPTX:CPTU // start the test
    // save the test report to the PC and rename it
Repeater AB test procedure via the command line interface

    CPTX:DUTT 3 // set up for repeater AB test
    CPTX:CPTR 20 // configure 3A-01 test
    CPTX:CPTU // start the test
    // save the test report to the PC and rename it

    // repeat above steps for test numbers 21 to 28

    CPTX:CPTR 29 // configure 3B-05 test
    CPTX:CPTU // start the test
    // save the test report to the PC and rename it
Repeater C test procedure via the command line interface

    CPTX:DUTT 2 // set up for repeater C test
    CPTX:CPTR 30 // configure 3C1-01 test
    CPTX:CPTU // start the test
    // save the test report to the PC and rename it

    // repeat above steps for test numbers 31 to 44

    CPTX:CPTR 45 // configure 3C2-09 test
    CPTX:CPTU // start the test
    // save the test report to the PC and rename it

Note: The release files can only be applied to the 882EA generator (not the 882CA/882C/881C, 882D, or 882E/881E generators).



Software Table

The Release 2.10.5 is comprised of the following software components.

Software Component Version Change Status
Boot Rom 01.04.11 no change
vxWorks 20.1884605 new with 2.10.5
Gateware 1 405C,1029,7252007 new with 2.10.5
Gateware 2 450A,1,6202007 no change
Gateware 3 450B,1,6132007 no change
Java JRE required 1.5.06 no change



Upgrade Instructions

You can upgrade your generator using the procedures in the Users Guide. Please note that you will have to verify that your current generator is operating with Release 2.10.0 or 2.10.2. To verify that your current generator is operating with one of these releases, enter the following command:

verf?

Release 2.10.0 should return the following information:

20.1884600, 01.04.11

Release 2.10.2 should return the following information:

20.1884602, 01.04.11

After verifying that the generator is operating with Release 2.10.0 or 2.10.2, upgrade your generator using the procedures below:

  1. Download the 2.10.5 Zip file to your PC and unzip the file. There will be a System and a Library directory.
  2. Upgrade the generator using the procedures at this link.
  3. After power cycling the generator, verify that the generator is running the current 2.10.5 version of the vxWorks file by entering the following command:

    verf?
    20.1884605, 01.04.11


  4. Verify that the vxWorks file is: 20.1884605.
  5. Verify that the BootRom version is: 01.04.11.
  6. Verify that the generator is running the current 2.10.5 version of the gateware files by entering the following command:

    verg?
    405C,1029,7252007:450A,1,6202007:450B,1,6132007


  7. The first listing is the Lattice FPGA gateware. The next two listings are the transmitter gateware files for the DVI and the HDMI output. The gatewares are separated by colons. Verify that the transmitter gateware files are: 450A,1,6202007:450B,1,6132007.


Other Release Notes

You can obtain Release Notes for past releases as follows:

882EA Release 2.10.2