 |
Known Issue #1: The ability to corrupt bits in the
CEC cores is broken. This affects CECT 9.5-1 of the CEC Compliance
Test. This will be fixed in a future release.
Known Issue #2: HDCP Compliance Test applet on the
web interface is broken. HDCP Compliance Test can be run via
front panel or command line.
The following features are included in the Release 2.10.2.
- EDID Compliance Test now supports HDMI CTS 1.3b.
The following anomalies have been corrected in the Release
2.10.2.
- Fixed the steady state and logic one levels on the Rx
DDC. Rx daughter card now functions properly as long as
resistors R19, R20, R58, and R59 are removed from the daughter
card. If you have an 882EA with release 2.10.0, please click
here for information on how to get the hardware modification
done on your Rx daughter card.
- General HDCP Compliance Test improvements.
- Fixed some problems with HdcpProd and HDCP query.
HDCP Compliance Test notes (only applicable to 882EA with
HDCP Compliance Test option)
Test coverage limitations of the HDCP Compliance Test
- Test steps TT01 and TT02 of Test Section 1A-01 are not tested because the application does not have visability into the control bits of the TMDS data islands. Therefore a Pass result on the test report only applies to the tests within that section that are executed.
- Test steps TT01 and TT02 of Test Section 1B-01 are not tested because the application does not have visability into the control bits of the TMDS data islands. Therefore a Pass result on the test report only applies to the tests within that section that are executed.
- Test Section 1A-08 is not tested because this test requires a DVD with a list of KSVs that have been revoked.
- Test Section 1A-09 cannot be tested because the application cannot collect the TMDS video.
- Test Section 2C-01 test S302 cannot be tested because the application cannot currently process Pj. Therefore a Pass result on the test report only applies to the tests within that section that are executed.
- Test Section 2C-04 cannot be tested because the application cannot collect TMDS data.
Applicable to all tests
- do not use CMD Terminal tool in the generator's web interface
to run the Compliance Test via command line; use Telnet
(Ethernet) or HyperTerminal (serial) instead
- try running the tests in batch mode by using the front
panel interface
- if there is a problem, try running the tests one by one
via the command line (see the repeater AB test procedure
below as an example and refer to the user
guide)
- running the tests using release 2.6.1 (20.1883610 firmware)
saves the test reports in /tffs0/library/reports/
- running the tests using a beta firmware (20.1883611 and
later) or release 2.10.2 (20.1884602 firmware) saves the
test reports in /card0/library/reports if there is a PCMCIA
card inserted in the 882CA/882EA (/tffs0/library/reports/
if no PCMCIA is inserted)
Applicable to the sink tests
- you can monitor the transactions on Tx port 1 using the ACA
Applicable to the source tests
- you can monitor the transactions if you have another 882C/882CA/882E/882EA
and a TPA-ACA-3R
- connect the output of the source device to TPA-ACA-3R's
DUT 1, the input 1 of 882CA/EA (test equipment) to TPA-ACA-3R's
DUT 2, and output 1 of a different 882C/882CA/882E/882EA
(ACA monitoring device) to TPA-ACA-3R's OUT 1
- use the ACA on the 882C/882CA/882E/882EA that is not the
test equipment to monitor the transactions on Tx port 1
Applicable to the repeater AB tests
- use a separate source device and connect it to the upstream port of the repeater
- use the 882CA/EA's Rx 1 port as the sink device and connect
it to the downstream port of the repeater
- you can monitor either downstream or upstream transactions
(not both at the same time) if you have another 882C/882CA/882E/882EA
and a TPA-ACA-3R
- in order to monitor the downstream transactions, connect
the downstream of the repeater device to TPA-ACA-3R's DUT
1, the input 1 of 882CA/882EA (test equipment) to TPA-ACA-3R's
DUT 2, and output 1 of a different 882C/882CA/882E/882EA
(ACA monitoring device) to TPA-ACA-3R's OUT 1
- in order to monitor the upstream transactions, connect
the output of the source device to TPA-ACA-3R's DUT 1, the
upstream of the repeater device to TPA-ACA-3R's DUT 2, and
output 1 of a different 882C/882CA/882E/882EA (ACA monitoring
device) to TPA-ACA-3R's OUT 1
- use the ACA on the 882C/882CA/882E/882EA that is not the
test equipment to monitor the transactions on Tx port 1
Applicable to the repeater C tests
- connect the 882CA/EA's Tx 1 port to the upstream port
of the repeater
- connect the 882CA/EA's Rx 1 port to the downstream port
of the repeater
- make sure HdcpProd image is not selected on the 882CA/882EA
before running the tests
- you can monitor the downstream transactions on Rx port 1 using the ACA
- you can monitor the upstream transactions on Tx port 1 using the ACA
Sink test procedure via the command line interface
CPTX:DUTT 0 // set up for source test
CPTX:CPTR 16 // configure 2C-1 test
CPTX:CPTU // start the test
// save the test report to the PC and rename it
// repeat above steps for test numbers 17 and 18
CPTX:CPTR 19 // configure 2C-4 test
CPTX:CPTU // start the test
// save the test report to the PC and rename it
Source test procedure via the command line interface
CPTX:DUTT 1 // set up for source test
CPTX:CPTR 1 // configure 1A-01 test
CPTX:CPTU // start the test
// save the test report to the PC and rename it
// repeat above steps for test numbers 2 to 14
CPTX:CPTR 15 // configure 1B_06 test
CPTX:CPTU // start the test
// save the test report to the PC and rename it
Repeater AB test procedure via the command line interface
CPTX:DUTT 3 // set up for repeater AB test
CPTX:CPTR 20 // configure 3A-01 test
CPTX:CPTU // start the test
// save the test report to the PC and rename it
// repeat above steps for test numbers 21 to 28
CPTX:CPTR 29 // configure 3B-05 test
CPTX:CPTU // start the test
// save the test report to the PC and rename it
Repeater C test procedure via the command line interface
CPTX:DUTT 2 // set up for repeater C test
CPTX:CPTR 30 // configure 3C1-01 test
CPTX:CPTU // start the test
// save the test report to the PC and rename it
// repeat above steps for test numbers 31 to 44
CPTX:CPTR 45 // configure 3C2-09 test
CPTX:CPTU // start the test
// save the test report to the PC and rename it
Note: The release files can only be applied
to the 882EA generator (not the 882CA/882C/881C, 882D, or
882E/881E generators).
The Release 2.10.2 is comprised of the following software components.
| Software Component |
Version |
Change Status |
| Boot Rom |
01.04.11 |
no change |
| vxWorks |
20.1884602 |
new with 2.10.2 |
| Gateware 1 |
405C,1029,7192007 |
new with 2.10.2 |
| Gateware 2 |
450A,1,6202007 |
new with 2.10.2 |
| Gateware 3 |
450B,1,6132007 |
new with 2.10.2 |
Java JRE required |
1.5.06 |
no change |
You can upgrade your generator using the procedures in the Users
Guide. Please note that you will have to verify that your
current generator is operating with Release 2.10.0. To verify
that your current generator is operating with one of these
releases, enter the following command:
verf?
Release 2.10.0 should return the following information:
20.1884600, 01.04.11
After verifying that the generator is operating with Release
2.10.0, upgrade your generator using the procedures below:
- Download the 2.10.2 Zip file to your PC and unzip the
file. There will be a System and a Library directory.
- Upgrade the generator using the procedures at this link.
- After power cycling the generator, verify that the generator
is running the current 2.10.2 version of the vxWorks file
by entering the following command:
verf?
20.1884602, 01.04.11
- Verify that the vxWorks file is: 20.1884602.
- Verify that the BootRom version is: 01.04.11.
- Verify that the generator is running the current 2.10.2
version of the gateware files by entering the following
command:
verg?
405C,1029,7192007:450A,1,6202007:450B,1,6132007
- The first listing is the Lattice FPGA gateware. The next
two listings are the transmitter gateware files for the
DVI and the HDMI output. The gatewares are separated by
colons. Verify that the transmitter gateware files are:
450A,1,6202007:450B,1,6132007.
|
 |